Description
The Bedside Monitor YR02173 is a cutting-edge monitoring solution designed to provide comprehensive patient information through its 12.1-inch touch screen with intuitive navigation. It supports critical parameters such as 5-lead ECG, RESP, and a single-channel temperature reading, along with advanced features like EtCO2 monitoring, AG analysis, BIS, and much more. This device is equipped with options for dual-IBP, thermal recording, and various mounting accessories, ensuring adaptability for a wide array of clinical environments.
Price in the Market
For medical institutions seeking to acquire the Bedside Monitor YR02173, it is essential to consider its market value, which ranges approximately from $2000 to $2800 USD. This range reflects the advanced functionalities and reliability offered by this high-performance unit, making it a valuable investment for enhancing patient care.
Frequently Asked Questions
- How does the Bedside Monitor YR02173 maintain accuracy?
It ensures precision with ±1% or ±1 bpm accuracy across its measurements, providing reliable data for informed decision-making.
- What is the display size of the monitor?
The monitor features a 12.1-inch touch screen, allowing for clear visibility and easy interaction.
- Is it compatible with wall mounting?
Yes, the monitor comes with options for both trolley and wall mounts.
Advantages and Disadvantages
Advantages: This monitor offers a comprehensive set of features, including dual-IBP and multiple mounting options, ideal for critical care settings. The touch screen interface enhances usability, and its connectivity to advanced modules like BIS and Nellcor/Masimo SpO2 ensures broad functionality.
Disadvantages: The initial investment may be higher compared to simpler models, but the enhanced capabilities justify the cost.
Product Usage in the Field
The Bedside Monitor YR02173 is primarily used in intensive care units and surgical wards, where continuous, accurate patient monitoring is critical. Its extensive parameter support and adaptability make it suitable for high-stakes environments where quick adjustments and reliable data are crucial for patient outcomes.
Recommendations
To maximize the monitor’s longevity, regular maintenance checks and software updates are recommended. Ensure all connections are secure and inspect sensors regularly to prevent data inaccuracies. Engage in the training provided by Kalstein Plus to fully leverage its capabilities.
Features
- 12.1 inch touch screen for easy navigation
- Supports 5-lead ECG, RESP, Single Channel Temp, COMEN SpO2, NIBP, HR
- Advanced features like dual-IBP, EtCO2, AG, ICG, C.O., BIS
- Nellcor/Masimo SpO2, 3/12-lead ECG, Thermal Recorder capabilities
- Comes with Trolley and Wall mount for versatile installation
Technical Specifications
Model | YR02173 | |
ECG
|
Lead mode | 5 Leads(RA,LA,LL,RL,V) |
Connection method | I, II, III,AVR,AVL,AVF,V | |
Lead Mode | 3 Leads (RA,LA,LL) | |
Connection method | I, II,III | |
Waveform | 1ch | |
Gain | 1.25, 2.5, 5.0, 10, 20, 40mm/mV, Auto. | |
HR Range | Adult: 15~300bpm | |
Neo/Ped | 15~350bpm | |
Accuracy | ±1% or ±1bpm | |
Resolution | 1bpm | |
Sensitivity | >200(uVp-p) | |
Differential input impedance | ≥5MΩ | |
ST segment Monitoring Range | ||
Measure Range | -2.0~+2.0mV | |
Measure Accuracy | -0.8mV~+0.8mV | |
Heart Rate
|
Measuring range | 15~300bpm |
Accuracy | ±1bpm | |
Resolution | 1bpm | |
Sensitivity | >200UVp-p | |
Respiration
|
Method | Impedance between R-F(RA-LL) |
Differential Input Impedance | >2.5MΩ | |
Measuring Impedance Range | 0.3~3.0 MΩ | |
Base line Impedance Range | 200~4000Ω | |
Bandwidth | 0.1~2.5Hz | |
Respiration Rate | Adult: 2~120BrPM | |
Neo/Ped | 0~150BrPM | |
Resolution | 1 BrPM | |
Accuracy | ±2 BrPM | |
Alarm | 10~40S | |
NIBP
|
Method | Oscillometric |
Mode | Manual/Auto/STAT | |
Measuring Interval in AUTO Mode | 1,2,3,4,5,10,15,30,60,90,120,180,240,480(Min) | |
Measuring Period in STAT Mode | 5 Min | |
Pulse Rate Range | 40~240bpm | |
Alarm Type | SYS, DIA, MEAN | |
Resolution | ||
Pressure | 1mmHg | |
Accuracy | Maximum Mean error: 5mmHg | |
Maximum Standard deviation | 8mmHg | |
SPO2
|
Display | Number of SPO2,waveform of capacity,Pulse |
Measuring range | 0~100% | |
Resolution | 1% | |
Accuracy | 60~100% (±2%) | |
Alarm system | Audio-video alarm | |
Pulse rate range | 0~300bpm | |
TEMP
|
Channel | 2 |
Measuring and Alarm Range | 0~50ºC | |
Resolution | 0.1ºC | |
Accuracy | 0.1ºC | |
Actualization interval | About 1 Second. | |
Average Time Constant | <10 Second. |